Humidity |
Resolution |
0.03% RH |
Repeatability |
±0.1% RH |
Accuracy & Interchangeability |
±2% RH (over range of 10% to 90% RH) |
Non-linearity |
<1% RH (after linearity compensation math)
|
Range |
0-100% RH |
Response Time |
4 seconds (slowly moving air, no cap) to within 67% of final value |
Hysterisis |
±1% RH |
Long Term Stability |
<1% RH per year typical |
Temperature |
Resolution |
±0.01% °C |
Repeatability |
±0.1 °C |
Accuracy |
±0.5 °C from 0°C to 40°C, <1°C from -20°C to +65°C |
Range |
-40°C to +123°C |
Response Time |
5 to 30 seconds to within 67% of final value |